62
9.
Remove the central screw of the ceramic terminal connector. Support the
terminal with one hand to keep it in place.
10.
Gently pull the ceramic terminal
and
the heating element at the same time,
horizontally toward the front of the machine.
11.
Discard the damaged heating element, and get the new one.
12.
Insert the leads of the new element in the matching holes of the ceramic
connector. (Do not tighten the terminal screws yet.)
13.
Gently insert the new element and the connector at the same time. Fasten the
connector firmly in place.
14.
Now tighten the connector screws on the element side, while looking into the
terminal holes on the other side, and making sure that the element leads are
correctly placed and squeezed by the terminal screws.
15.
Now insert the wires in the terminal on the opposite side of the element, and
tighten then in the terminal.
16.
Tighten the stopper screw corresponding to the re-installed heating element.
17.
Repeat steps 6 to 15, until you have replaced all elements that needed to be
changed.
18.
Now, ensure that all heating elements are fully inserted.
19.
Tighten the two screws of the element stabilizer clamp, while pushing
downwards on that clamp.
20.
Re-install the LED/safety guard.
21.
Re-install the top/back panel.
Top/back
panel removal
In order to access the furnace and the rocking system, the top/back panel of the
instrument must be removed.
IMPORTANT
: Always disconnect the instrument from the power outlet when
working with panels removed. 240 Volts inside! Risk of electrocution!
1.
Disconnect the instrument from the wall outlet.
2.
Partially open the safety shield, to just disengage its locking mechanism.
(Refer to page 16,
Operation of the safety shield
, to unlock the safety shield
when power is turned off.)