36
The three rows (1, 2 and 3 on the right-hand side) allow the user to optionally set
three distinct fan speeds (in percent) and corresponding durations. This permits
special “quenching” operations; most fusion users do not need to use those and
will simply set zero-durations for two of those sub-steps and 100% fan power for
a few minutes in the last one, as in this example:
Note on the tilting angles during cooling
Typically, the tilting angle of “Cooling 1” will be the same as the pouring, and the
tilting angle of “Cooling 2” will be ninety degrees (90°) to prevent residual drops
from sliding on the outside wall of the crucible, or fall onto the cooling bead. This
will give some time for the flux to pour completely out of the mold during “Cooling
1”, while partially straightening the crucibles when the blower starts at the
beginning of “Cooling 2”.
Unlocking the
advanced
mode
Before being allowed to manage fusion program and edit parameters,
one must enter the correct password. To do so, click on the padlock
icon/button.