93
Fehler erkennen und
beheben
Detección y subsanación de
errores
Recognizing and remedying
errors
6. 2 Error indications for EASY
monitoring
Error indications are displayed in three stages
(green, yellow and red) when problems occur.
When working with a split neutral electrode, the
following errors may occur:
EASY monitoring
Cause
Indication
Remedy
Lights up yellow
Significant increase in resistance
Depending on the indication, there may
be heating under the neutral electrode
–
Stopping the application is not necessary.
Check the positioning of the neutral electrode.
Switches from green to
continuous red
A significant problem occurrs when the
unipolar current is activated.
An acoustic signal
sounds.
An error message
appears on the display.
Check the neutral electrode and neutral electrode cable,
see section EASY monitoring, page 28.
e
Check the neutral electrode cable for proper connection
and external damage.
Loosened electrode
An acoustic signal
sounds.
A warning message
appears on the display
e
Correct the positioning of the neutral electrode. If the
error persists, replace the neutral electrode.