Installation
4-2
96296081
EN /
BA
4. 3 Installation and Comissioning
The device can be operated free-standing, in a rack or as a wall-mounted solution in eit-
her a vertical or horizontal position.
4.3.1 Minimum requirements
• Ambient conditions for operation as described in section 1.7.
• Consideration of the technical data as described in section 3.3.
• Consideration of the free area dimensions based on the graphic below to guarantee
that the device is properly ventilated.
1
note: nothing
may be mounted to the left and right of the device in the defined free
areas.
85,9
310
85,9
20
375
20
Device Enclosure
Device front panel
Defined open area
Max. temperatur 35° C
Air
inlet
Air
outlet