Recommended separation distances between portable and mobile RF communications equipment and the 9619 NB / 9626 NB(-2)
The
9619NB/9626NB(-2)/9627NB(-2)
is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or user of the
9619NB/9626NB(-2)/9627NB(-2)
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
9619NB/9626NB(-2)/9627NB(-2)
as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
[W]
Separation distance d
[
m
]
according to frequency of transmitter
150 kHz to < 80 MHz
D
80 MHz to < 800 MHz
D
800 MHz to 2.5 GHz
F
0.01
0.2
0.2
0.4
0.1
0.4
0.4
0.8
1
1.2
1.2
2.4
10
3.7
3.7
7.4
100
11.7
11.7
23.4
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the
|
transmitter manufacturer.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects