T
echnical manual OT-600-060/1
Page 7
Monitoring of the mattress pressure
The system will alarm if the pressure deviates from set value. If a compressor failure (or big leak)
prevents the set pressure from being reached then the fans and heating will not start.
If the electronic pressure control fails to stop the compressor then a mechanical over pressure valve will
prevent the pressure from exceeding 75mb. A sudden pressure drop (puncture or loose hose) will
create an alarm and stop the system.
Monitoring of the fans
The system uses two high quality silent fans that periodically changes the air direction in order to
achieve an even temperature distribution under the patient. In case one fan stops the system will alarm
and enter into “safe mode” (See Error code 009). Change of temperature must be done after careful
consideration of the fact that the air is constantly going in the same direction (“hot-spot” hazard).
Monitoring of the temperature regulation
The electronic temperature sensors (one in each hose connection at the Main Unit) will alarm if the air
temperature deviates from selected value. Mean air temperature and peak temperature values are
monitored. The system is also equipped with 2 capillary thermostats that will cut the heating if the
electronics should fail. As a third independent passive patient protection the mattress is equipped with
a temperature sensitive weak point that will puncture the mattress if the temperature should pass
critical values.
7. Technical Data
Voltage
220 - 240 VAC 50 - 60 Hz
Power consumption
150 VA (Average under normal conditions), 700 VA (Peak value)
Inflation time
About 90 sec with 200 cm Mattress.
Warming up time
About 4 minutes from room temperature (18-20°C).
Measurements
Main Unit: 310 x 295 x 370 mm
Length of Air Hoses: 1500 mm
Hand Unit: 115 x 145 x 27 mm
Length of cable: 3000 mm
Weight
Main Unit: About 27 kg incl. wheelbase. Hand Unit about 800 g
Transport packing in plywood box: About 34 kg
Temperature selection
20°C - 42°C, step of 1°C Pre- 1°C
Pressure selection
0 – 60 mb. Step of 5 mb. Pre- 5mb.
Mechanical overpressure valve release at 75 mb
Function supervision
Audio and visual alert. Error codes displayed on the hand unit with a code key on
the back of the hand controller.
Environmental data
+15°C to +35°C during operation, less than 90% RH
0°C to +40°C during storage, non-condensing.
- 40°C to +50°C (max 56°C/30 min) during transport, non-condensing.
Modifications
Any modifications on the control unit, hand unit, hoses and the mattress will void
KANMED’s responsibilities totally and are not allowed without the written consent
of KANMED.
Creating Systems
Any person connecting the WarmCloud to any other systems mains socket may
have created a “system” according to IEC 60601-1, and must be qualified to
evaluate any consequences that may harm staff, patient and equipment.