4. Cleaning, Checks before use and Maintenance
Cleaning
Unplug the power cord before cleaning. Clean the Main Unit and Hand Controller as per the hospital
instructions for medical technical equipment.
The Mattress is a disposable item and shall not be reused.
Checks before use
Check that all parts are clean and in good condition. Check that the unit performs a self test at start. If
Error Codes appear and cannot be eliminated with the suggestions given in section 10 then they should
be notified to Technical staff unless they are of type A (See Trouble Shooting and error codes section
10).
Note
: If the Main Unit or Hand Controller has received mechanical damage, it should always be
checked and correct system function verified by technical staff.
Maintenance
KANMED WarmCloud does not require any special maintenance apart from what is described in the
technical section of this manual.
The soft part of the air hose Mattress connection needs to be changed if worn. If the air hoses are worn
or leaking they must also be changed.
5. Accessories and spare parts
Always check with your dealer or
www.KANMED.se
for the latest information.
Article nr.
Product
Comment
OT-600-001
KANMED WarmCloud
Complete system with standard accessories.
OT-600-002
Hand Controller
With rubber edges and 3m cable
OT-600-004
Accessory strap
Transport band with elastic strap.
OT-600-220
Warm-Air Mattress,
Standard
Single-Use Mattress, Length 200 cm, Width 50 cm. Delivered in box
of 10 pcs
OT-600-222
Warm-Air Mattress, Wide
Single-Use Mattress with ties, Length 200 cm, Width 60 cm.
Delivered in box of 10 pcs
OT-600-211
Warm-Air Mattress, Small
For “leg-up” positions and for Children
Single-Use Mattress with ties, Length 100 cm, Width 50 cm.
Delivered in box of 10 pcs
OT-600-701 Hose
set
complete
Consumable
OT-600-702
Hose seal (2 pcs)
Consumable
OT-600-704
Fuses (2 pcs)
T 6.3 A L 250V, 5x20 mm
OT-600-705
Cable 3 m for Hand Unit
Spare part, complete with connectors
OT-600-070 Manual
English
Manual.
OT-600-060
Manual
Service manual English.
6. Technical information
Safety systems
KANMED WarmCloud OT600 is equipped with several safety functions to prevent dangerous situations.
If a fault is detected an alarm will sound and an error code will appear on the Hand Controller.
Depending on the seriousness of the error the system will continue to function normally, enter a safe
mode or shut down. See “Error Codes” later.
An alarm of less seriousness will, if still persisting, reappear after 10 minutes.
A serious fault in the warming system will shut down the warming but the pressure regulation of the
mattress will continue to function.
The green lamp over pressure and temperature displays indicates the status of the regulating circuits.
A steady glow indicates that set values have been achieved. A blinking lamp indicates that the unit is
working on reaching set values.