T
echnical manual OT-600-060/1
Page 11
10. SERVICE INFORMATION
Modifications
Any modifications on the control unit, control unit electronics or the air mattress will void KANMED’s
responsibilities totally and are not allowed without the written consent of KANMED
For a general description of the WarmCloud and how to verify that it is functioning properly we refer to
the user manual. In case you do not have one it can be downloaded from internet in several languages.
Se
www.KANMED.se
This service manual describes the basic functions of the WarmCloud and how do determine faults and
do smaller repairs.
KANMED’s service policy states that only KANMED authorised service points may carry our repairs on
the main PCB.
Instead we ask you to make full use of our PCB exchange program because this ensures the delivery of
a tested and calibrated replacement PCB with 3 months warranty.
KEY COMPONENTS
1) Pressure sensor
2) Temperature sensor (x2)
3) Thermostat (x2)
4) Heating element (x2)
5) Circulation fan (x2)
6) Electrical over pressure valve
7) Compressor
8) Mechanical over pressure valve
The main PCB interfaces and controls the key components. No calibration or PCB repairs are to be done
(PCB, and key component replacement only).
2
K
A
B
1
4
3
5
7
6
+
-
+
-
Hand Unit
8
Main Unit
Alarm