50
PRODUCT SPECIFICATION
UNIT
Temp Measurement
Resolution Range
Accuracy
Flue Temperature
0.1
o
(C/F)
0-1100°C
32-2140°F
* Use high temperature probe for gases
>600°C/1112°F
1.0
o
C +0.3% of
reading
Inlet Temperature
0.1
o
(C/F)
0-600°C
0-999°F
1.0
o
C +0.3% of
reading
Gas Measurement
*1
Resolution Range
Overrange
Reading
Accuracy
Oxygen (0
2
):
0.01%
0-25%
30%
-
-0.1% +0.2%
Carbon monoxide (CO):
(standard: H
compensated)
1ppm
2000
4000
<100ppm
>100ppm <2000ppm
>2000ppm <4000ppm
+/-5ppm
+/-5% of reading
+/-10% reading
Nitric oxide (NO):
(high range0)
1ppm
1000
5000
<100ppm
>100ppm <1000ppm
>1000ppm <5000ppm
+/-5ppm
+/-5% of reading
+/-10% reading
Nitric oxide (NO)
(low range)
1ppm
100
300
<100ppm
>100ppm <300ppm
+/-5ppm
+/-10% of reading
Nitrogen dioxide (NO
2
):
1ppm
100
1000
<100ppm
>100ppm <1000ppm
+/-5ppm
+/-10% of reading
Sulphur dioxide (SO2)
(low range):
1ppm
100
500
<100ppm
>100ppm <500ppm
+/-5ppm
+/-10% of reading
Sulphur dioxide (SO
2
)
(high range):
1ppm
2000
5000
<100ppm
>100ppm <2000ppm
>2000ppm <5000ppm
+/-5ppm
+/-5% of reading
+/-10% reading
Hydrogen sulphide
(H
2
S):
1ppm
200
300
<100ppm
>100ppm <200ppm
>200ppm <300ppm
+/-5ppm
+/-5% reading
+/-10% of reading
Gas Measurement
*1
Resolution Range
Accuracy
Pressure
0.01mbar
0-150 mbar
+ 0.5% Full Scale
Carbon dioxide (CO
2
)*2
0.1%
0 – Fuel Value
+ 0.3%
Efficiency
*2
0.1%
0-100%
+ 1%
*1
using dry test gases at STP
*2
calculated
NB: all ppm reading can be displayed in mg/m
3
and can be normalised