O P E R AT I O N C O N T I N U E D
5.
Type "zerocal" on either command box and click on "Send ASCII." The OOD will average a few readings and
calibration results will be displayed along with the confirmation message. FIG. 4-7.
6.
The OOD is now calibrated and will read current pipeline conditions as 0 ppm oil.
SETTING THE ALARM SETPOINT
Please follow the steps below once RealTerm has been installed and configured as per instructions on the
"Communication Software Configuration" section on pages 20-22.
1.
Ensure power to the OOD is turned off. To access the boards, use a 7/16" wrench to remove the (6) hex screws
on the Explosion Proof electronics enclosure and remove the cover. Connect your PC to the OOD sensor via the
supplied RS232 serial cable and USB to serial converter as per wiring diagram on page 14 of this manual.
2.
Turn on power to the OOD and launch RealTerm.
3.
Go to the "Send" tab and type "OODmaster" (case sensitive) in either command box. Click on "Send ASCII." There
will be a return message: "System has entered Admin mode!" FIG. 4-8.