11
OODMANUAL 0720
KAM CONTROLS, INC.
REMOVING THE OOD™ SENSOR
1.
To remove the OOD™ sensor, first turn off power
and disconnect all electrical connections to the
OOD™ enclosure.
2.
Make sure that the line pressure is below 100 psi.
Loosen the Hex Nuts, using a 3/4" wrench, on the
top of the Locking Collar one quarter to one half
turn.
3.
Slowly and with caution, use a 3/8" Allen wrench
tp loosen the Socket Cap Screws on the Locking
Collar. In cases where line pressure cannot be
reduced below 100 psi, a KAM IT Insertion Tool
must be used to safely remove the unit.
CAUTION: Once the Socket Cap Screws have been
loosened, the OOD™ shaft may push out from the
line. If pressure in the line is above 100 psi, it may
do so with enough force to cause bodily injury or
damage to the instrument.
4.
Slide the OOD™ sensor upward until it stops and
the probe rests inside the seal housing. FIG. 3-13.
5.
Next, close the Full-opening Ball Valve tightly.
The OOD™ sensor may now be unbolted or
unscrewed from the system.
FIG. 3-13
Socket Cap Screws
I N S TA L L AT I O N C O N T I N U E D
Hex Nuts