49
Note
Possible errors: 2 identical calibration points or temperature
coefficient larger than 5.5%
- The instrument indicates an error.
✱
After pressing the
key or EXIT, the selected calibration
procedure is shown.
8.6 Calibrating the relative cell constant
General
Because of manufacturing variations, each conductivity cell has a
real cell constant which deviates slightly from the ideal nominal cell
constant. The real cell constant may drift due to dirt accumulating
on the cell.
This deviation of the real from the nominal cell constant is reflected
by the relative cell constant. The relative cell constant is determined
through calibration.