7 Setting ranges
0 — 1 µS to 0 — 200 mS, depending on the cell constant
Unit
0 = S/cm
1 = mho/cm
Cell
constant
Range
0.01
1
/
cm
1
0 – 1.000 µS/cm
0 – 1.000 µmho/cm
1
2
0 – 2.00 µS/cm
0 – 2.00 µmho/cm
1
3
0 – 5.00 µS/cm
0 – 5.00 µmho/cm
1
4
0 – 20.00 µS/cm
0 – 20.00 µmho/cm
2
0.1
1
/
cm
5
0 – 5.00 µS/cm
0 – 5.00 µmho/cm
1
6
0 – 20.00 µS/cm
0 – 20.00 µmho/cm
1
7
0 – 200.0 µS/cm
0 – 200.0 µmho/cm
2
8
0 – 1000 µS/cm
0 – 1000 µmho/cm
3
1
1
/
cm
9
0 – 500.0 µS/cm
0 – 500.0 µmho/cm
1
10
0 – 1000 µS/cm
0 – 1000 µmho/cm
3
11
0 – 2.00 mS/cm
0 – 2.00 mmho/cm
2
12
0 – 10.00 mS/cm
0 – 10.00 mmho/cm
3, 4
13
0 – 20.00 mS/cm
0 – 20.00 mmho/cm
2
14
0 – 100.0 mS/cm
0 – 100.0 mmho/cm
3, 4
3
1
/
cm
15
0 – 30.00 mS/cm
0 – 30.00 mmho/cm
3, 4
10
1
/
cm
16
0 – 100.0 mS/cm
0 – 100.0 mmho/cm
3, 4
17
0 – 200.0 mS/cm
0 – 200.0 mmho/cm
3
The deviations from the characteristic listed below refer to
µS/cm or mS/cm
1
deviation from characteristic
≤
1%
2
deviation from characteristic
≤
1.5%
3
deviation from characteristic
≤
2%
4
Above a temperature of
≥
85°C and a temperature coefficient T
K
> 2.2%/°C, the deviations from the characteristic may be larger.