44
8.4 Calibrating the temperature coefficient using
automatic temperature measurement
Note
During calibration, the reference temperature and the working
temperature can be approached in any sequence.
✱
Immerse the conductivity cell and the temperature sensor in the
medium to be measured.
The currently measured temperature is displayed.
WORK.T shows you that the working temperature that will be used
later on has to be entered.
Note
The reference and working temperatures must differ by at least 5°C.
✱
Enter the working temperature (e.g. 40.0 for 40.0°C), continue
with .
✱
Temper the medium to be measured to the reference
temperature.
- The LC display shows the temperature in the upper line, and the
uncompensated conductivity in the line below.
The conductivity value will be accepted automatically on reaching
the reference temperature,
or
blinks
P