8 Technical Data
35
Direct current
NiCrSi-NiSi „N“
DIN EN 60584
-100 … +1300°C
0.4%
Pt10Rh-Pt „S“
DIN EN 60584
0 … +1768°C
0.4%
Pt13Rh-Pt „R“
DIN EN 60584
0 … +1768°C
0.4%
Pt30Rh-Pt6Rh „B“DIN EN 60584
300 … 1820°C
0.4%
W3Re-W25Re „D“
0 ... 2495°C
0.4%
Cold junction
Pt 100 internal
Cold junction accuracy
± 1K
Measuring range
210 ms, 420ms with double thermocouples (C112=1)
Input filter
digital filter, 2st priority; filter constant can be set from 0 … 100s
Particularities
can also be programmed in °F
1.The accuracy values refer to the maximum measuring range.
Measuring range
Accuracy
0 … 20mA, voltage drop < 2V
4 … 20mA, voltage drop < 2V
0.2%
0 … 10V, input resistance > 100 k
Ω
2 … 10V, input resistance > 100 k
Ω
0.1%
Scaling
can be freely programmed within the limits
Measuring range
210 ms
Input filter
digital filter, 2st priority; filter constant can be set from 0 … 100s