163
7 Configuration
7.18.5
Standard operation
When standard operation is activated, the measurement data is recorded at the configured memory cy-
cle unless one of the following operating modes is active:
•
Event operation
•
Time operation
Standard operation has the lowest priority compared with time and event operation.
Device dialog box
Parameter
Parameter
Selection/settings
Description
Memory status
On
The data is saved if this operating mode
is active.
Off
The data is
not
saved if this operating
mode is active.
Memory values
Average value
The average value across the configured
memory cycle is calculated and saved.
Current value
The current value at the end of the mem-
ory cycle is stored.
If the signal returns an error value during
the memory cycle, even if this is only for
a short time, the value is stored.
Min. value
The minimum value within the configured
memory cycle is stored.
Max. value
The maximum value within the config-
ured memory cycle is stored.
Min./Max. value
The minimum and maximum values with-
in the configured memory cycle are
stored.
?
?