Safety installations, warning functions
32
8.4. Limitation of the operating temperature range
Factory settings:
t
H
igh
S
etpoint: 200 °C
t
L
ow
S
etpoint: -50 °C
The limitation of the operating temperature range effects the temperature
setting under the menu called up with the
key. It is possible to
adjust only working temperatures that lie within the limit range set here.
Existing settings for Setpoint 1, 2, 3 and also for > t High < and > t Low <
(see page 31) are automatically defered within the limit range.
1.
Press the key
until the menu item
or
is displayed.
2.
Press enter
to indicate the adjusted value on the MULTI-
DISPLAY (LED) (digit blinks).
3.
Set value:
Use the cursor keys
to move left or right on the display until
the numeral you wish to change blinks.
Use the increase/decrease arrows
to change the selected
numeral (-, 0, 1, 2, 3, ... 9).
4.
Press enter
to store the selected value.
Press
to update the display immediately.