Recirculating Cooler
25
Recommended bath fluids:
Bath
fluid
Temperature
range
soft/decalcified water
5 °C to 80 °C
See website for list of recommended bath fluids.
Contact:
see page 6
Caution:
Fire or other dangers when using bath fluids that are not recommended:
Please contact JULABO before using other than recommended bath liquids.
JULABO assumes no liability for damage caused by the selection of an unsuitable
bath fluid.
Unsuitable bath fluids are fluids which, e. g.,
x
are highly viscous
(much higher than 30 mm
2
x s
-1
at the respective working temperature)
x
have a low viscosity and have creep characteristics
x
have corrosive characteristics or
x
tend
to
crack.
No liability for use of other bath fluids!
ATTENTION:
The maximum permissible viscosity is 30 mm
2
/s
5.3. Filling
Notice:
¾
Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the liquid.
Liquid level
x
Connect the tubing from the external system to the pump connectors
and check for leaks.
x
Check to make sure that the drain port (25) is closed.
x
Remove the cap from the filling opening (9).
x
Fill the bath tank using a funnel while monitoring the filling level (6).
Filling of the external system.
x
Turn the mains switch (1) on
x
Press the key
for filling the cooling loop for the external system.
Make sure that air can evacuate from the system..
x
Check the filling level (6) and keep on filling the bath liquid using the
funnel.
x
After having finished the filling process, the liquid level should be
below “MAX”.
x
Close the filling opening.