Operating procedures
28
6.3. AUTOSTART ON / OFF
n
Keep depressed enter
and the start/stop key
o
and turn on the circulator with the mains power switch.
For a short while the MULTI-DISPLAY indicates the effective start mode:
Ö
AUTOSTART on.
Ö
AUTOSTART off.
NOTE:
The recirculating cooler has been configured and supplied by JULABO
according to N.A.M.U.R. recommendations. This means for the start
mode, that the unit must enter a safe operating state after a power
failure (non-automatic start mode). This safe operating state is indicated
by „OFF“ or „rOFF“, resp. on the MULTI-DISPLAY (LED). A complete
shutdown of the main functional elements such as heater and circulation
pump is effected simultaneously.
Should such a safety standard not be required, the AUTOSTART
function (automatic start mode) may be activated, thus allowing the start
of the circulator directly by pressing the mains power switch or using a
timer.
Warning:
For supervised or unsupervised operation with the AUTOSTART function, avoid any
hazardous situation to persons or property.
The recirculating cooler does no longer conform to N.A.M.U.R. recommendations.
Take care you fully observe the safety and warning functions of the circulator.