29
immediately during
rescue.
The defibrillator needs
repair.
Contact Jousing
Medical for service.
The iAED-S1 prompts
“Service required”,
“device is not ready for
use”, “begin CPR now”
during rescue.
The defibrillator needs
repair.
Perform CPR and
replace the device
immediately, then
contact Jousing Medical
for service after this
rescue.
Status indicator is solid
in red, and the iAED-S1
prompts “Battery low” in
battery insertion
self-test.
Battery is low.
Replace a new battery
immediately after
performing this rescue.
Or, replace a new
battery immediately.
Status indicator is solid
in red, and the iAED-S1
prompts “Replace
battery now” in battery
insertion self-test.
Battery is depleted.
Replace a new battery
immediately.
Status indicator is solid
in red, and the iAED-S1
prompts “Service
required” in battery
insertion self-test.
The defibrillator needs
repair.
Contact Jousing
Medical for service.
Status indicator is solid
in red, and the iAED-S1
prompts “abnormal
operatin
g temperature”
in battery insertion
self-test.
The device has placed
beyond the temperature
range of 0
℃
~ 50
℃
.
You can also perform
rescue if it needs of
rescue. Or, place the
device at room
temperature, and wait
for 10 minutes, then
reinstall the battery, it
will run a self-test again
to eliminate equipment
failure alarm.
Status indicator flashes
in red, and the iAED-S1
is chirping in standby
mode.
Battery is low.
Replace a new battery.
The defibrillator needs
repair.
Contact Jousing
Medical for service.
The device has placed
beyond the temperature
range of 0
℃
~ 50
℃
.
Place the device at
room temperature, and
wait for 10 minutes,
then reinstall the
battery, It will run a
self-test again to
eliminate equipment
failure alarm.