12
device, as shown in figure 4.
Warning:
Use only the pads approved by Jousing Medical, the pads
specified in the 1.2 Device Appearance.
Figure 4 Pads Connector Insertion
3.4 (Optional) Replacing the Battery
Scenarios
Replace the battery when the device prompts “Replace the battery” or “Battery
low”.
Warning:
⚫
Do not remove the battery from the device optionally.
⚫
Battery lifetime will be shortened if the defibrillator turns on frequently.
Procedure
Press up the spring button on the right side of the device and pull out the
battery, as shown in Figure 5 a). Remove the battery from the battery holder as
shown in figure 5 b). After removing the old battery, wait at least 30 seconds
before reinstalling a new battery.