25
socket and pads connector.
⚫
Check the device for damage.
⚫
Check the scratches or signs of damage, especially near the connector
and the connector socket. If there is any scratch or sign of damage,
contact Jousing Medical for service.
6.2 Cleaning the Device Surface
Scenarios
Periodically clean the iAED-S1 and keep it away from dust or dirt.
Please use the household detergents of soapy water or chlorine bleach
(10ml/liter). Do not use strong solvents such as isopropyl alcohol or acetone to
clean the iAED-S1.
Follow the guidelines when cleaning the device:
⚫
Remove the battery.
⚫
Use a soft cloth dampened in the cleaning agent to clean the device surface.
Do not immerse the iAED-S1 in water or allow fluids to spill onto it.
⚫
If necessary, clean the excess agents on device with water or a dry cloth.
⚫
Ensure the iAED-S1 is completely dry before reinstalling its accessories.
6.3 Storage Environment
The device should be stored near the emergency equipment (such as first aid
kits) with suitable temperature. Keep it away from moisture and dust.
⚫
The suggested storage temperature range is 5
℃
~ 35
℃
.
⚫
The device should be stored in relative humidity of 5% ~ 95%
(non-condensing).
⚫
The pads only has 24 hours lifetime at the temperature: -30
℃
~ 0
℃
, 50
℃
~
65
℃
, and the pads only has 4 months lifetime at the temperature: 35
℃