21
4.3.7 No Shock Required
If the iAED-S1 determines that
no shock is required, the device will prompt “No
shock advised”, The user will be prompted to begin CPR, refer to 4.3.8 CPR.
4.3.8 CPR
Procedure
1.
Compress the patient’s chest when the device prompts “da, da” beat.
2.
Make artificial respiration to the patient when the device prompts “Breath”.
Details
Rescue cardiac arrest patients with artificial respiration and chest
compression.
When rhythm analysis has determined no shock is required or a shock has
been delivered, the device will prompt “It is now safe to touch the patient, begin
CPR, da, da...” user should compress the patient’s chest following the beat.
When the device prompts “Breath”, user should make artificial respiration to
the patient.
When the device prompts “Stop CPR”, it will restart to analyze the heart rhythm,
no one should touch the patient during that time.
The CPR coaching protocol can be changed. Only the authorized person can
change the CPR protocol by using specific software. Refer to Annex A for more
information.
4.3.9 Terminate Operation
For any reason you want to turn off the defibrillator during the operation, press
the ON/OFF button and hold for at least 3 seconds to turn it into standby mode.