28
in red, and the iAED-S1
prompts “replace battery
now” in rescue.
immediately.
Status indicator is solid
in red, and the iAED-S1
prompts “Service
required” during rescue.
The defibrillator needs
repair.
Contact Jousing
Medical for service.
The iAED-S1 prompts
“Poor pad contact to
patient, press pads
firmly”, “Check
connector”.
Pads are not correctly
applied to the patient.
Make sure the pads
have been removed
from the liner and place
pads on patient to the
right place.
Pads are not making
good contact with the
patient.
Dry the patient’s chest
and shave or clip any
excessive chest hair,
and make sure the pads
are not touch the
patient’s clothing.
Connector is not
inserted.
Make sure pads
connector
is inserted correctly
Pads, pad cable or pad
connector may be
damaged.
Replace the pads.
Pads socket may be
damaged.
Contact Jousing
Medical for service.
The iAED-S1 prompts
“stop motion”,” analysis
interrupted” during
rescue.
Patient is moving.
Check patient’s
breathing
Someone is
touching/moving the
patient or doing CPR to
the patient.
Stop touching/moving or
doing CPR to the
patient
Vehicle is moving.
Stop the vehicle during
analysis, if possible.
The iAED-S1 prompts
“shock button not
pressed, shock
cancelled” during
rescue.
Shock button is not
pressed within 30
seconds.
Push shock button
within 30
seconds
The shock cannot be
delivered during rescue.
The pads may be
damaged.
Replace the pads.
Battery is depleted.
Replace a new battery.
Device turns off
Battery is depleted.
Replace a new battery