5. After Using the iAED-S1
5.1 Replacing Pads
After rescuing, press the ON/OFF button to turn the device into standby mode.
Change the pads into new ones: Insert the pads plug into the pads socket( see
3.3 Checking the Pads Connection.) and put the pads package in the carrying
bag.
Warning:
The disposable pads must be discarded and replaced after use.
5.2 Checking the Battery Capacity
Remove the battery and wait for at least 30 seconds to reinsert and run
self-test.
⚫
If the device prompts “Self-test completed, battery low” or “Self-test
completed, replace battery now”. Replace the battery immediately and
ensure that the new battery is within the expired date.
⚫
If the battery insertion self-test result is normal, no action needs to be
taken.