3.2 Battery Insertion Self-Test
Procedure
Wait for battery insertion self-test completion.
If there are some prompts during the self-test, be sure to follow the prompts
and press corresponding buttons and let the self-test run until it completes.
Verification
⚫
If the status indicator flashes green from fast to slow and the voice prompts
“self-test completed device normal”, it means the device is ready for use.
⚫
If the status indicator flashes red, the device will prompt:
◼
“Self-test completed, replace battery now” User should replace a new
battery immediately after performing this rescue if a rescue is needed,
or replace a new battery immediately.
◼
“Self-test completed, battery low”. User should replace the battery
immediately.
◼
“Self-test completed, temperature abnormal”. You can also perform
rescue if needed. Or place the device in room temperature and wait for
10 minutes, then reinstall the battery to eliminate equipment failure
alarm.
◼
“Self-test completed, service required”. That means error is detected
and user should contact Jousing Medical for service.
3.3 Checking the Pads Connection
Procedure
1. Make sure the pads package is intact and within expired date. Replace the
pads if it is expired or the pads package is damaged.
2. Insert the pads connector into the device if
it’s not fully inserted into the