2.0 EGOLD-1000 Installation

 

 

2.1 Important Installation Points 

 



Check that you have selected the proper Tank Probe length to suit your particular interceptor tank 
requirements. Tank Probe length is based on inside wall height of the interceptor tank. The proper 
Tank Probe length should be 2 inches less then the inside wall height   

 

 

 

For example: An interceptor tank with an inside wall height of 15 inches will require Tank Probe 
assembly part number MA47-13. An interceptor tank with an inside wall height of 20 inches will 
require Tank Probe assembly part number MA47-18.  

 
 

Note: For inside tank wall heights greater than 22 inches consult JOSAM for the proper Tank 
Probe assembly part number.  

 



Prior to mounting the Tank Probe assembly in the interceptor tank, insure that the tank probe 
location chosen will not interfere with any baffle elements in the interceptor tank when installed. 

 



Installation will require drilling of the interceptor tank wall. 

Note: Care should be taken when 

operating power tools around the interceptor tank due to the presence of water.

 

 



It is important to carefully follow the calibration procedure outlined below, in order to insure 
proper grease level detection. 

 



The wall mount power module that powers the Control Unit must be directly plugged into 
an outlet. Do not use an extension cord. If electric service is required, a licensed and 
certified electrician must install it. 

 

 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for EGOLD-1000

Page 1: ...000 Electronic Grease Level Detector Installation and Operating Procedures Josam Company 2501 South Front Street Philadelphia PA 19148 215 339 5370 WARNING Cancer and Reproductive Harm www p65warnings...

Page 2: ...EGOLD 1000 CONTROLS AND INDICATORS 9 4 0 EGOLD 1000 CALIBRATION 11 5 0 EGOLD 1000 OPERATION 13 6 0 EGOLD 1000 MAINTENANCE 15 Figures FIGURE 1 1 ILLUSTRATES THE E LECTRONIC GREASE LEVEL DETECTION SYST...

Page 3: ...nted Control Unit and the Tank Probe assembly The Control Unit GLS housing connectors control switches and indicators are water resistant for splash or spray conditions The Tank Probe assembly is a se...

Page 4: ...EG0LD 1000 Electronic Grease Level Detector Installation and Operating Procedures __________________________________________________________________________________________ Page 4 Figure 1 1 MODEM...

Page 5: ...20 inches will require Tank Probe assembly part number MA47 18 Note For inside tank wall heights greater than 22 inches consult JOSAM for the proper Tank Probe assembly part number Prior to mounting t...

Page 6: ...rceptor and within 5 feet of an 115VAC wall outlet 2 Place the Control Unit on the wall and mark the two 2 mounting hole locations Drill 3 16 holes to accept the mounting hardware 3 Use the mounting h...

Page 7: ...th the 20 screw and wing nut provided Bracket B should be free to slide up and down on bracket A Attach the probe to the underside of bracket A with the 20 screw provided Slide the probe and bracket B...

Page 8: ...el Detector Installation and Operating Procedures __________________________________________________________________________________________ Page 8 Figure 2 1 STATIC WATER LEVEL OUTLET BRACKET A BRACK...

Page 9: ...stem is operating properly in the grease level detection mode Calibration LED When ON this LED indicates that the EGOLD 1000 Electronic Grease Level Detector system is in the grease level calibration...

Page 10: ...or Installation and Operating Procedures __________________________________________________________________________________________ Page 10 2 5mm POWER JACK CENTER POSITIVE MOUNT USING 8 HARDWARE PROV...

Page 11: ...eptor tank wall that is approximately 1 3 of the distance from the bottom of the outlet pipe to the bottom of the tank Note When grease fills the interceptor tank down to this level the alarm will be...

Page 12: ...interceptor tank with water until water flows out of the interceptor tank outlet pipe 10 Press the Cal Set Button on the Control Unit The two 2 Green LEDs will begin flashing After several minutes the...

Page 13: ...he Yellow LED indicates the grease volume is approaching the level required for cleaning the grease interceptor tank The Red LED indicates the grease volume has reached the level requiring the grease...

Page 14: ...operating errors as follows The Operating LED will be OFF The Red Level LED will be flashing The audible Alarm will sound intermittently short beeps once every ten seconds System errors include Loss o...

Page 15: ...be cleaned every time the grease is emptied from the interceptor tank As a minimum the tank probe should be cleaned every 6 months Tank probe cleaning can be accomplished by wiping the probe with a s...

Page 16: ...onic Grease Level Detector Installation and Operating Procedures __________________________________________________________________________________________ Page 16 Appendix I Tank Probe Assembly Mount...

Page 17: ...00 Electronic Grease Level Detector Installation and Operating Procedures __________________________________________________________________________________________ Page 17 Appendix II Optional Alarm...

Page 18: ...hat is within 15 feet of the Control Unit within 5 feet of an 115VAC wall outlet and within 5 feet of a telephone jack to place the Modem Unit provided 2 Plug the jack connector of the modem cable pro...

Reviews: