Mounting/Wiring

Installation

Remote sensor mounting 

procedure 

1. Cut an opening in the wall of the 

isolation room to receive the supplied 

single-gang “old work” low voltage 

mounting bracket (Figure 1) for the 

remote sensor module. Nominal 

dimensions for the cutout are 3.65” H x 

2.15” W. Drill a 7/16” hole through the 

opposite wall for the flow tube. 

2. Install the single-gang low voltage 

mounting bracket in the cutout. Route 

a length of supplied flow tube through 

the mounting bracket and through the 

7/16” hole in the opposite wall.  

 

Separate the backplate from the 

touchscreen display enclosure by 

turning the set screw at the bottom of 

the display enclosure fully clockwise 

to release the backplate. Pull the 

backplate bottom out slightly to 

clear the bottom edge of the display 

enclosure and slide the backplate down 

away from the tabs at the top of the 

enclosure. Disconnect either end of the 

gray harness between the backplate 

and display module by pushing down 

on the locking tab, and then sliding the 

connector out. Disconnect the interface 

cable from the display module by 

removing the terminal block from the 

4 pin connector CN3 on the display 

module PC board. 

3. Confirm that the 4-conductor 

interface cable between the remote 

sensor module and the touchscreen 

display module has its red and black 

conductors securely attached to the 

+V and GND terminals, respectively. 

Route the end of the interface cable 

with the 3 and 4 pin connector down 

the wall to the anticipated location of 

the touchscreen display module.

4. Using a suitable length of 2 conductor 

20 AWG cable, connect the output of 

the universal isolated power supply 

module to the power input at the 

remote sensor module. The output 

of this power supply module is non-

polarized, so the two red wire may be 

wired to power input terminals in either 

orientation. Refer to page 12-13 for 

power wiring details.   

5. The FMS-1655M Monitor has a 

factory-calibrated analog output and is 

available as voltage at the Vo terminal. 

The designated output (Vo) can be 

used to control a damper actuator to 

maintain a user set pressure setpoint 

when the FMS-1655M Monitor is used 

in a controller application.  

6. Attach the pluggable connector to the 

mating header on the remote sensor 

module. Attach the end of the flow tube 

to the barbed fitting at back of sensor, 

and then secure the remote sensor 

module to the mounting bracket with 

two supplied 6-32 x ¾” screws. 

7. Install the louvered cover plate onto 

the installed sensor module using 

two supplied machine screws. On the 

opposite side of the wall (corridor side), 

attach the end of the flow tube to the 

barbed fitting of the flow tube mounting 

plate. 

8. Press the mounting plate into place, 

allowing the excess tube length to 

go into the wall space. Secure the 

mounting plate with the supplied 

screws and wall anchors. 

9. Install a louvered cover plate 

onto the mounting plate using two 

supplied machine screws.

Fig 1. Single-gang, low-voltage mounting 

bracket.

 

 

Summary of Contents for FMS-1655M

Page 1: ...l apparatus meets all the requirements of the Canadian Interference Causing Equipment Regulations Cet appareil numérique de la Classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada IMPORTANT The FMS 1655M Room Pressure Monitor must be wired to 24 VAC only Wiring the unit to 110 VAC will cause serious damange and void the warranty Risk of Property Damage Ensure th...

Page 2: ...rning Setpoint Positive Isolation Low Alarm Setpoint Negative Isolation High Alarm Setpoint Negative Isolation High Warning Setpoint Negative Isolation Low Warning Setpoint Negative Isolation Low Alarm Setpoint Audible Alert Operating Mode audible or slient Delay Time Base seconds or minutes Delay Setting 0 60 Alarm Quiet Period Starting Hour 0 23 Alarm Quiet Period Ending Hour 0 23 Engineering Un...

Page 3: ... Display mounting procedure surface mount 10 Flush thin mount mounting procedure 11 Installation detail view 12 Basic programming 14 Main display screen 14 Configuring Room Pressure Monitor 14 Setting up alarm limits 14 Selecting displayed units 15 Configuring display options 15 Adding password security 16 Cleaning the display 17 Alarm setpoints 18 Setup menu tree 19 ...

Page 4: ...r Depth 18 bit or 262K colors Backlight Color White Luminous Intensity min 2500 cd m2 FMS 1655M Monitor Surface Mount Enclosure 3 W x 5 H x 1 13 D FMS 1655M Monitor Thin Mount Enclosure 5 W x 8 H x 0 75 D External Remote Sensor Housing 2 3 W x 4 H x 2 7 D Stainless Steel Cover Plate for Flow Tube 2 7 W x 4 5 H x 0 2 D Stainless Steel Cover Plate for Remote Sensor 2 7 W x 4 5 H x 0 2 D Surface Moun...

Page 5: ...655M LIT 12013119 5 Part Number Guide UNIT FMS Flow Monitor Station FMS 01 Standard remote sensor REMOTE SENSOR S Surface mount T Thin flush mount STYLE SERIES BRAND LB Johnson Controls 1655M BACnet Monitor ...

Page 6: ...ansformer be split between actuator and controller Doing so will damage the actuator the transformer the controller or all units A single 120 24V 30Va transformer is required for the controller and a separate 120 24V 20Va transformer is required for the actuator This equipment contains electrostatic sensitive components To prevent possible damage take precautions to prevent electrostatic discharge...

Page 7: ...g in this compartment Note This product should be installed with the manufacturer provided isolated power supply and connected to an electrical circuit protected by a minimum 20A circuit breaker This circuit breaker should be mounted in an approved electrical enclosure located separately but in close proximity to this product 240 VAC 50 60 Hz Blue 1 Amp Slow Blow Fuse Transformer 50 60 Hz ...

Page 8: ...MS 1655M Monitor provides a single digital input that may be used for monitoring a door switch The configuration of the door switch input is configured for normally closed operation and is active high triggered A SPDT magnetic door switch type is recommended for use with function The user may set up multiple multi level passwords to prevent unauthorized or casual access to the FMS 1655M Monitor co...

Page 9: ... respectively Route the end of the interface cable with the 3 and 4 pin connector down the wall to the anticipated location of the touchscreen display module 4 Using a suitable length of 2 conductor 20 AWG cable connect the output of the universal isolated power supply module to the power input at the remote sensor module The output of this power supply module is non polarized so the two red wire ...

Page 10: ...m all 4 conductors of the interface cable are securely attached at the terminal plugs Re connect gray harness between backplate and display module 6 Reconnect the gray harness to the backplate circuit board and attach the display enclosure to the backplate by inserting the tabs at the top of the display into the corresponding slots at the top of the backplate Secure the display enclosure to the ba...

Page 11: ...wall and discard Be sure to brush off any drywall dust or remnants from the inside surface of the opening to ensure proper adhesion of the retrofit mounting plate 4 Remove the paper backing from the two adhesive strips on the retrofit mounting plate and insert it into the cut opening of the drywall The retrofit plate should be oriented such that the corner notch is located at the lower left corner...

Page 12: ...all to seal unit penetration Flow Tube When Flow Tube mounting plate is located directly opposite the sensor flow tubing must be cut as short as possible to prevent kinks To Johnson Controls isolation transformer Optional Analog Output Damper Control To optional door switch Wall Section cutaway view Terminal For connection of transmitter to FMS 1655M Monitor Orange wall bracket to be installed fir...

Page 13: ...ss steel plate and wall to seal unit penetration Terminal For connection of transmitter to FMS 1655M Monitor Flow Tube When Flow Tube mounting plate is located directly opposite the sensor flow tubing must be cut as short as possible to prevent kinks To FMS 1655M Monitor monitor Wall Section cutaway view Rotating clamps secure assembly to wall board Optional Analog Output Damper Control To Johnson...

Page 14: ...open if door switch is enabled or 2 current differential pressure has drifted outside of the allowable limits of the desired setpoint and are in the caution range A red background indicates that the current differential pressure has reached a critical condition and is outside of the allowable limits of the desired setpoint The FMS 1655M Monitor incorporates a full color touchscreen and includes an...

Page 15: ...the audible alerts to be suppressed between the specified hours every day thereby eliminating the potential for nuisance alarms Hospitals may take advantage of this feature to minimize nuisance alarms during non visiting hours in patient rooms Selecting displayed units The FMS 1655M Monitor displays differential pressure readings in one of two units inches of water column in WC or Pascals Pa Touch...

Page 16: ...e FMS 1655M Monitor is able to communicate 11 points of information to the building management system See page 18 for list of BACnet points The network settings are easily configured by selecting the Network Setup option from the Unit Setup menu As an alternate means of remote monitoring a 4 20mA signal representing the room DP is available for connecting to a third party controller on the BMS It ...

Page 17: ...pping the current room name text brings up an alphanumeric keyboard to quickly change the name of the monitored room Tap anywhere else on the screen enters the Main Setup Menu if no password is stored Otherwise a password must be entered before the Main Setup Menu can be accessed Tapping the units brings up engineering units selection popup to quickly change pressure measurement units Tapping Time...

Page 18: ...FMS 1655M LIT 12013119 18 Alarm setpoints ...

Page 19: ...FMS 1655M LIT 12013119 19 Setup menu tree ...

Page 20: ...FMS 1655M LIT 12013119 20 ...

Page 21: ...scription Read or Write Analog Inputs AI 1 Analog Input 1 default Differential Pressure Read Only Binary Inputs BI 1 Digital Input 1 Door Switch Read Only Analog Outputs AO 1 Analog Output PID 1 OUT Read Only Analog Values AV 1 Setpoint Differential Pressure Setpoint Read Write AV 2 Low Pressure Alarm Setpoint Read Write AV 3 Low Pressure Warning Setpoint Read Write AV 4 High Pressure Warning Setp...

Page 22: ...FMS 1655M LIT 12013119 22 FMS 1655M Monitor wiring guide ...

Page 23: ...respective owners Copyright 2019 Johnson Controls All rights reserved Any unauthorized use or copying is strictly prohibited European Single Point of Contact JOHNSON CONTROLS WESTENDHOF 3 45143 ESSEN GERMANY NA SA Single Point of Contact JOHNSON CONTROLS 507 E MICHIGAN ST MILWAUKEE WI 53202 USA APAC Single Point of Contact JOHNSON CONTROLS C O CONTROLS PRODUCT MANAGEMENT NO 22 BLOCK D NEW DISTRICT...

Reviews: