9

Joerns® Support Surface

FlapCair Model

© 2013 Joerns Healthcare • 6110066 RevD • 13-2631

Position Sensor

Patient 

Position Sensor 

(Fowler boost) will be 

illuminated when the head of the bead is elevated by 

more than 30 degrees.

Modes

Autofirm

Autofirm 

mode provides maximum air inflation 

designed to assist both patients and caregivers 

during patient transfer and treatment. The unit 

will automatically return to the mode it was in 

prior to 

Autofirm 

(either 

Pulsate 

or 

Therapy

) in 

approximately 12 minutes.

Therapy Modes: Therapy and Alternating 

 

Pulsate

Pulsate 

will slightly decrease the pressure 

in all cushions every 30 seconds, which some 

patients find comfortable.

 

Therapy

: The unit starts in the 

Therapy 

mode, 

which is the standard low air loss therapy. 

Comfort Adjust

The 

Comfort Adjust

 function is located on the right 

side of the control panel. The FlapCair can be 

customized to meet individual patient needs within a 

therapeutic window. See “Setup” section for further 

information. This function will not work in 

Autofirm

 

mode. 

Lockout

This feature is to prevent any unauthorized 

changes to the patient settings. To unlock and make 

adjustments to the settings press both up and down 

comfort arrows at the same time to disengage the 

Lockout 

function. The 

Lockout 

function will return in 

approximately five minutes.

Note:

 The unit is designed to lock out all the 

adjustment controls after the patient has been 

positioned correctly. In approximately five minutes 

after the last button push the power on light begins 

to flash indicating 

Lockout 

is enabled.

Additional Features

Warning: A possible fire hazard exists

. This 

product is suitable for use with oxygen administering 

equipment of the nasal, mask, or half bed-length, 

tent-type only. To prevent personal injury or 

equipment damage, ensure that the oxygen tent 

does not extend below the mattress.

CPR

The hose connection at the control unit is marked 

CPR

. Disconnect the hose from the control unit. 

Deflation times will vary based on patient weight and 

profile. To resume therapy, reattach the hose to the 

control unit.

Transport

To transport the patient in bed, turn the control unit 

off. Unplug the power cord from the outlet. Do not 

disconnect the hose connection at the control unit. 

The lower support cell will prevent the patient from 

bottoming out for up to 12 hours.

Power Failure

The FlapCair control unit has an advanced power 

failure notification alarm and Comfort Adjust setting 

memory capability. Upon power failure the control unit 

will flash a red Alarm light on the keypad and sound an 

audible signal every two (2) seconds for up to twenty 

(20) minutes. The power interruption feature keeps the 

current settings in memory based on the frequency of 

use and level of battery charge. The Alarm Reset will 

turn off audible and visual alarm.

9

Figure 5

Summary of Contents for FlapCair

Page 1: ...User Service Manual Joerns Support Surface FlapCair To avoid injury read user s manual before using...

Page 2: ...s Do not use in the presence of smoking materials or open flame Air flowing through the air mattress will support combustion 2 Danger To reduce the risk of shock adhere to the following instructions F...

Page 3: ...nufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Joern...

Page 4: ...al Features 9 10 CPR Transport Power Failure CairRails Optional Accessories Troubleshooting 10 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Control Nursing Procedures...

Page 5: ...gle caregiver installation featuring rapid inflation in just five minutes or less The user friendly controls allow for easy adjustment of patient comfort Additionally low friction and low shear materi...

Page 6: ...avoid unwanted or accidental adjustments Specially designed enclosure for low noise and durability Integrated swing out hanging brackets for fixing to most bed types 6 Therapy Mattress System Feature...

Page 7: ...e therapy pad is not already on the mattress place it on the mattress Attach the zippered cover to the mattress surround Hang the control unit on the foot of the bed facing away from the bed Attach th...

Page 8: ...at are mobile will need to be monitored to ensure they remain in the correct position to float the wound Operation Warning For important precautions please see page two Caution The patient s head shou...

Page 9: ...ve minutes Note The unit is designed to lock out all the adjustment controls after the patient has been positioned correctly In approximately five minutes after the last button push the power on light...

Page 10: ...ge Ready Valve for instant inflation and deflation Transfer friendly deflate for ease of assisted transfer or when bolsters are not required Unique contoured design allows ease of ingress egress while...

Page 11: ...incontinence barrier pad to absorb the excess moisture In the event of incontinence or excess drainage on the therapy pad you should wipe off the excess fluid from the bed surface 11 Safety Informatio...

Page 12: ...f there is no visible soilage with possible body fluids we recommend that you clean the mattress system with a mild detergent and warm water If disinfection is desired you may use a combination cleans...

Page 13: ...or Assistance If you have any questions or require service on a Joerns product please call Joerns Healthcare at 800 826 0270 Support Surface Gently roll up the support surface expelling any residual a...

Page 14: ...andling and or cleaning Failure to use in the manner indicated in the FlapCair user manual Any modification repair or alteration done to the FlapCair that was not authorized in writing by Joerns Healt...

Reviews: