9
Joerns® Support Surface
FlapCair Model
© 2013 Joerns Healthcare • 6110066 RevD • 13-2631
Position Sensor
Patient
Position Sensor
(Fowler boost) will be
illuminated when the head of the bead is elevated by
more than 30 degrees.
Modes
Autofirm
Autofirm
mode provides maximum air inflation
designed to assist both patients and caregivers
during patient transfer and treatment. The unit
will automatically return to the mode it was in
prior to
Autofirm
(either
Pulsate
or
Therapy
) in
approximately 12 minutes.
Therapy Modes: Therapy and Alternating
•
Pulsate
:
Pulsate
will slightly decrease the pressure
in all cushions every 30 seconds, which some
patients find comfortable.
•
Therapy
: The unit starts in the
Therapy
mode,
which is the standard low air loss therapy.
Comfort Adjust
The
Comfort Adjust
function is located on the right
side of the control panel. The FlapCair can be
customized to meet individual patient needs within a
therapeutic window. See “Setup” section for further
information. This function will not work in
Autofirm
mode.
Lockout
This feature is to prevent any unauthorized
changes to the patient settings. To unlock and make
adjustments to the settings press both up and down
comfort arrows at the same time to disengage the
Lockout
function. The
Lockout
function will return in
approximately five minutes.
Note:
The unit is designed to lock out all the
adjustment controls after the patient has been
positioned correctly. In approximately five minutes
after the last button push the power on light begins
to flash indicating
Lockout
is enabled.
Additional Features
Warning: A possible fire hazard exists
. This
product is suitable for use with oxygen administering
equipment of the nasal, mask, or half bed-length,
tent-type only. To prevent personal injury or
equipment damage, ensure that the oxygen tent
does not extend below the mattress.
CPR
The hose connection at the control unit is marked
CPR
. Disconnect the hose from the control unit.
Deflation times will vary based on patient weight and
profile. To resume therapy, reattach the hose to the
control unit.
Transport
To transport the patient in bed, turn the control unit
off. Unplug the power cord from the outlet. Do not
disconnect the hose connection at the control unit.
The lower support cell will prevent the patient from
bottoming out for up to 12 hours.
Power Failure
The FlapCair control unit has an advanced power
failure notification alarm and Comfort Adjust setting
memory capability. Upon power failure the control unit
will flash a red Alarm light on the keypad and sound an
audible signal every two (2) seconds for up to twenty
(20) minutes. The power interruption feature keeps the
current settings in memory based on the frequency of
use and level of battery charge. The Alarm Reset will
turn off audible and visual alarm.
9
Figure 5