Power Interruption (Brown out)

In the event of a brief power interruption, up to twenty 

(20) minutes, the Flapcair control unit will automatically 

restart and reset to the previous Comfort Adjust 

settings. The power interruption feature keeps the 

current settings in memory based on the frequency of 

use and level of battery charge.

Power Loss

In the event of a power failure in excess of twenty 

(20) minutes, the control unit’s stand by light will be 

illuminated after power restoration. Press the Power 

button and reset the preferred mode and comfort level 

as described in the “Operation” section. 

Safety cell

In the event of a power failure, the lower support 

cells should remain inflated for up to 12 hours as 

long as the control unit remains connected. In case 

of an extended power failure, transfer the patient to a 

hospital mattress or other surface.

CairRails

Integrated CairRails risk management air bolsters offer 

a bilateral side bolster solution designed to address 

healthcare’s growing concerns of liability in relation 

to patient falls and entrapment. CairRails are being 

recognized by some of the nations leading healthcare 

systems for improving their patient safety and risk 

management programs.

CairRails are recommended for patients requiring 

additional support during patient care and transfer. 

CairRails can help reduce costs while ensuring optimal 

clinical outcomes and increasing patient safety.

Note: 

When inflating CairRails, it is recommended that 

the control unit be in 

Autofirm 

mode to achieve optimal 

results.

Features and Benefits

 

A bilateral side air bolster solution which can 

enhance your facilities entrapment/risk management 

program.

 

Easy to engage 

Ready Valve 

for instant inflation and 

deflation. 

 

• Transfer friendly-deflate for ease of assisted transfer 

or when bolsters are not required.

 

Unique contoured design allows ease of ingress/

egress, while providing additional protection, comfort 

and supports patient compliance.

 

• Designed to fit on most key Joerns therapeutic 

support surfaces.

 

Promotes maximum independence by  allowing 

caregiver to decide when added protection is 

required. 

Note:

 CairRails are meant to provide a documentable 

and functional intervention for the risk management 

issues of falls and entrapment but in no way guarantee 

the prevention of falls or entrapment occurrences.

Optional Accessories

 

Additional therapy pads – available for purchase

Troubleshooting

Therapy Surface is Not Inflating

• 

Ensure the hose connection from the therapy 

mattress system (mattress) to the control unit is 

securely connected.

• 

Ensure that the control unit is plugged into  

an AC outlet.

• 

Ensure that the power is not on 

Standby.

 If on 

Standby

, press the 

Power

 button.

• 

Ensure that all air cells are connected to the internal 

mattress manifold.

Unable to Change Therapy Mode or Adjust 
Comfort Control

Make sure the

 Lockout

 function is disabled. To 

disable, press the up and down 

Comfort Adjust

 arrows 

simultaneously.

Nursing Procedures

Recommended Linen

Special linens are not necessary for the FlapCair. 

There is no need for a bottom sheet as the therapy 

pad should be covering the therapy cells at all times. 

The patient should never be lying directly on the 

therapy cells. Upon the patient specific needs the 

following linens may be utilized:

• 

Draw or slide sheet to aid in positioning and to 

further minimize friction and shearing.

• 

Incontinence barrier pad for patients incontinent of 

urine and/or stool, and patients with heavily draining 

wounds.

• 

Add top sheet, blanket and/or bedspread as needed 

for patient comfort.

• 

Keep the amount of padding between the patient 

and bed to a minimum for optimum performance.

10

Summary of Contents for FlapCair

Page 1: ...User Service Manual Joerns Support Surface FlapCair To avoid injury read user s manual before using...

Page 2: ...s Do not use in the presence of smoking materials or open flame Air flowing through the air mattress will support combustion 2 Danger To reduce the risk of shock adhere to the following instructions F...

Page 3: ...nufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Joern...

Page 4: ...al Features 9 10 CPR Transport Power Failure CairRails Optional Accessories Troubleshooting 10 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Control Nursing Procedures...

Page 5: ...gle caregiver installation featuring rapid inflation in just five minutes or less The user friendly controls allow for easy adjustment of patient comfort Additionally low friction and low shear materi...

Page 6: ...avoid unwanted or accidental adjustments Specially designed enclosure for low noise and durability Integrated swing out hanging brackets for fixing to most bed types 6 Therapy Mattress System Feature...

Page 7: ...e therapy pad is not already on the mattress place it on the mattress Attach the zippered cover to the mattress surround Hang the control unit on the foot of the bed facing away from the bed Attach th...

Page 8: ...at are mobile will need to be monitored to ensure they remain in the correct position to float the wound Operation Warning For important precautions please see page two Caution The patient s head shou...

Page 9: ...ve minutes Note The unit is designed to lock out all the adjustment controls after the patient has been positioned correctly In approximately five minutes after the last button push the power on light...

Page 10: ...ge Ready Valve for instant inflation and deflation Transfer friendly deflate for ease of assisted transfer or when bolsters are not required Unique contoured design allows ease of ingress egress while...

Page 11: ...incontinence barrier pad to absorb the excess moisture In the event of incontinence or excess drainage on the therapy pad you should wipe off the excess fluid from the bed surface 11 Safety Informatio...

Page 12: ...f there is no visible soilage with possible body fluids we recommend that you clean the mattress system with a mild detergent and warm water If disinfection is desired you may use a combination cleans...

Page 13: ...or Assistance If you have any questions or require service on a Joerns product please call Joerns Healthcare at 800 826 0270 Support Surface Gently roll up the support surface expelling any residual a...

Page 14: ...andling and or cleaning Failure to use in the manner indicated in the FlapCair user manual Any modification repair or alteration done to the FlapCair that was not authorized in writing by Joerns Healt...

Reviews: