To configure the
NAVIC
system for scanning using dual drive modules with a probe
holder frame, follow these steps
(see “Probe Holder Frame” on page 89)
:
1.
Connect power supply
(Fig. 37-2)
to user umbilical breakout
(Fig. 37-5b)
using
the power cable
(Fig. 37-9)
.
2.
Connect left
(Fig. 37-10)
and right
(Fig. 37-4)
drive modules to crawler umbilical
breakout
(Fig. 37-5a)
.
3.
Mount crawler umbilical breakout
(Fig. 37-5a)
to right drive module
(Fig. 37-4)
.
4.
Connect controller
(Fig. 37-3)
to user umbilical breakout
(Fig. 37-5b)
using the
controller cable
(Fig. 37-7)
.
NOTE:
The controller may also be connected directly (Fig. 37-8) to the crawler
umbilical breakout (Fig. 37-5a).
5.
Connect encoder cable
(Fig. 37-6)
from the instrument
(Fig. 37-1)
to the user
umbilical breakout
(Fig. 37-5b)
.
6.
Attach a configured probe holder frame
(Fig. 37-11)
(see “Probe Holder Frame” on
page 89)
to the crawler
(see “Swivel Mount” on page 44)
.
CAUTION!
Do not disconnect under load.
Shut off power before connection or disconnecting.
Permanent damage to electronics could occur.