BA-40100-0-V04 | Maintenance Instructions TOPAX DX | 35
Chapter 16: Menu structure and configuration
Menu 2.2: Alarms
mg/l
mV
pH
039 702 719 09
0
ESC
free chlorine
pH value
REDOX
temperature
next
setup
mg/l
alarms
2.2
main menu >> menu >> alarms
mg/l
mV
pH
039 702 719 09
0
ESC
comb. cl.
% salt
alarms
setup
2.2
main menu >> menu >> alarms >> next >> page (at more than 5 inputs)
The TOPAX DX monitors its inputs (see chapter 9). If an un-
wanted alteration appears on an input value, the alarm is set off.
The alarm-setting is configured in this menu.
Menu
Function
Required value
2.2.2
Free chlorine
..1.1
Maximum Alarm
OFF or ON + value
..1.
Minimum Alarm
OFF or ON + value
..1.3
Y-Alarm
OFF or ON + value
..1.4
Alarm delay time
OFF or ON + value
2.2.2
pH value
...1
Maximum Alarm
OFF or ON + value
...
Minimum Alarm
OFF or ON + value
...3
Y-Alarm
OFF or ON + value
...4
Alarm delay time
OFF or ON + value
2.2.3
Redox
..3.1
Maximum Alarm
OFF or ON + value
..3.
Minimum Alarm
OFF or ON + value
..3.4
Alarm delay time
OFF or ON + value
Menu
Function
Required value
2.2.4
Temperature
..4.1
Maximum Alarm
OFF or ON + value
..4.
Minimum Alarm
OFF or ON + value
..4.4
Alarm delay time
OFF or ON + value
2.2.5
Combinded chlorine
..5.1
Maximum Alarm
OFF or ON + value
..5.
Minimum Alarm
OFF or ON + value
..5.3
Y-Alarm
OFF or ON + value
..5.4
Alarm delay time
OFF or ON + value
2.2.6
Salt content
..6.1
Maximum Alarm
OFF or ON + value
..6.
Minimum Alarm
OFF or ON + value
..6.3
Y-Alarm
OFF or ON + value
..6.4
Alarm delay time
OFF or ON + value
Select one of the six inputs. The device goes to the current menu
.. … ..5 and shows the actual configuration of the input
alarm.
In the next step select for configuration one (of up to four) alarm
functions for each input:
Alarm
Alarm will be activated, when ...
Button
Maximum measuring
value
Input value > max. value
max Alarm
Minimum measuring
value
Input value < min. value
min Alarm
Y-Alarm
after a given time for the input of the
relevant control unit in the compu-
ter’s output service is > 95%
Y-Alarm
Alarm delay time
the alarm’s delay time has run out
Delay
Possible alarm functions of the six inputs.
Press on “OFF” to turn off the chosen alarm function. Confirm
twice with the “OK” key.
Press on “ON” to turn the chosen alarm function on.
Enter the requested value:
The first digit of the target value is highlighted in grey and can be
adjusted. Use the arrow keys to set up the value:
◀
/
▶
switch over between the individual positions
▲
/
▼
alter numerical values
„OK“
to save (press times)