Chapter 4 Performing Calibration
38
CellAdvisor JD720C User's Manual
Figure 9
e-Calibration connection diagram with an extension cable
JD723C/724C
JD725C/726C
Procedure
1
Connect an extension cable, if necessary, to the
RF Out / Reflection
port of the instrument.
2
Connect the
EZ-Cal
to the instrument's the
RF Out / Reflection
port or at the end of the extension
cable if used.
3
Connect the USB cable to the side of the
EZ-Cal
and then to the
USB Host
port to power the
e-calibration kit.
4
Press the
CAL
hard key.
The on-screen instruction appears to guide you through the e-calibration.
5
Do the following depending on your instrument model and its installed option:
–
JD723C, JD724C, and JD726C without 2-port transmission option: Tap the
Continue
button.
The calibration progress bar appears.
–
JD725C and JD726C with 2-port transmission option: Tap the
O-S-L
button.
The calibration progress bar appears.
6
Check the screen to see if the calibration status is now changed to
CAL ON (E)
.
Figure 10
e-Calibration on JD724C using EZ-Cal kit