Chapter 12 High Power CW Signal Generating
102
CellAdvisor JD720C User's Manual
1
To set the frequency of the signal, complete the following steps:
a
Tap the
Frequency
screen key.
b
Enter a value within the frequency range of your instrument by using the numeric keys,
and then press the
ENTER
hard key. You can also use the rotary knob.
2
To set the output level of the signal, complete following steps:
a
Tap the
Power Level
screen key.
b
Enter a value between
-30 dBm
and
10 dBm
by using the numeric keys, and then press the
ENTER
hard key. You can also use the rotary knob.
3
To start generating the set signal, toggle the
RF Source
screen key, and then select
On
.
4
To stop generating the signal, toggle the
RF Source
screen key, and then select
Off
.