Chapter 13 RF Power Testing
CellAdvisor JD720C Series User
’s Manual
107
Setting display
You can manually set minimum and maximum values of the measurement plot that vary
depending on the connected power sensor.
Procedure
1
Press the
SETUP
hard key and touch the
Display
tab.
The Display setup window appears.
2
Select the cell next to the
Minimum
for
Average
PS
,
Peak
PS
,
VSWR
, or
Pulse
as needed.
3
Enter a value by using the numeric keys or the rotary knob.
4
Select the cell next to the
Maximum
for
Average
PS
,
Peak
PS
,
VSWR
, or
Pulse
as needed.
5
Enter a value by using the numeric keys or the rotary knob.
6
Tap the
Apply
button in the lower-right corner of the setup screen to save the change(s).
Figure 65
Display setup window for RF power measurements
Making a measurement
Once the connected power sensor is initialized and you have configured parameters, the
instrument is now ready to make a measurement.
Procedure
1
Connect the power sensor plugged in to your instrument with the RF output port of BTS as shown
in
for a terminating power sensor or
for a directional power sensor.
2
Your measurement result is displayed on the screen in real time.