24
Initial start up
When power is first connected, the display shows a menu page asking for the language to be
selected followed by calibration. Other menu items cannot be selected at this stage. This is
for your own safety, as the device would measure meaningless values if calibration had not
been successfully carried out. Once these stages have been completed, any other selections
can be made.
If the value overview (chapter 4) appears instead of the “Language” selection, the device
has already been used (demonstration etc.). In this case, carry out by all means a calibration.
Simply touch the icon “settings” and select the menu item “Calibration”. Confirm with OK.
Continue as described below.
Preparation
Please use the calibration liquids and cuvettes supplied together with the JBL PROFLORA
pH Cal. Fill each calibration cuvette up to the 10 ml mark, filling one cuvette with
pH 7.00 buffer solution, one with pH 4.00 buffer solution and one with deionised water (JBL-
Dest). The buffer solutions are coloured with an indicator in order to prevent confusion. For
more stability, place the calibration cuvettes in the 3 large holder spaces on the calibration
stand. Unscrew the screw fitting on the storage tube of the sensor by about one turn and
pull out the sensor. Put the sensor together with the temperature sensor in the calibration
cuvette with distilled water, move it around a little and leave the two sensors in the
water. Because the pH measurement and the calibration are temperature-dependent, it is
important that both sensors are immersed in the calibration liquid.
Language selection
Press the arrow symbols to select the required language and press OK. The following
language options are displayed: DE/GB/F/NL/IT/DK/ES/PT/RO
Summary of Contents for PROFLORA pH Control Touch
Page 2: ...1 a...
Page 3: ...2...
Page 4: ......
Page 8: ...4 Kalibrierung...
Page 29: ...25 DE UK FR NL IT ES PT SE CZ PL RU RO TR Calibration...
Page 50: ...46 talonnage...
Page 71: ...67 Kalibratie DE UK FR NL IT ES PT SE CZ PL RU RO TR...
Page 92: ...88 Calibratura...
Page 113: ...109 DE UK FR NL IT ES PT SE CZ PL RU RO TR Calibraci n...
Page 134: ...130 Calibra o...
Page 155: ...151 DE UK FR NL IT ES PT SE CZ PL RU RO TR Kalibrering...
Page 195: ...191 DE UK FR NL IT ES PT SE CZ PL RU RO TR Kalibracja...
Page 216: ...212...
Page 217: ...213 7 00 4 00 calibration o k 4 2 3 7 1 7 13 30 CO DE UK FR NL IT ES PT SE CZ PL RU RO TR...
Page 218: ...214 6 5 4 3 2 1 7 8 9 10 11 12 13 1 2 3 4 5 Auto pH KH 6 7 8 9 10 11 CO 12 13 Auto pH...
Page 226: ...222 OK 0 10 2 00 0 05 1 00 CO 0 50 CO 20 C t 30 C 4...
Page 228: ...224 OK locked unlocked OK locked OK OK 7422 4 software version Software Version OK...
Page 230: ...226 4 5 2 sensor defect DELTAV OFFSET DELTAV 35 7 00 OFFSET pH 7 00 4 00 7 00...
Page 232: ...228 24 JBL 2 3 JBL JBL pH Sensor 24 JBL JBL PROFLORA pH...
Page 238: ...234 Calibrare...
Page 259: ...255 DE UK FR NL IT ES PT SE CZ PL RU RO TR Calibration Kalibrasyon...
Page 277: ...273...