Jazzy Series
57
Corrective Maintenance
If the battery condition meter does not light up when you turn on the power:
Check the harness connections. Make sure they are tight.
Check the circuit breaker. Reset it if necessary.
Check the battery connections.
If the above conditions prove normal, you can load test the batteries with a battery load tester. These
testers are available at automotive parts stores. Disconnect both batteries before load testing and follow
the directions that come with the load tester. If either one of the batteries fails the load test, replace both
of them. If your power chair still does not power up, contact your authorised provider.