Operation
15
RM/TM-2040GE Series
Figure 6.
Bayer 2
3.1.3
Full Scan Area 2x2 Binning
TM-2040GE has 2x2 binning of the full scan area. In full scan 2x2 binning mode, pixel (i, j) includes all the
information of pixel (2i-1, 2j-1), (2i-1, 2j), (2i, 2j-1) and (2i, 2j) in normal full scan mode (where i=1 2, ..., 800;
j=1,2, ...., 600). In this mode vertical binning makes frame transfer faster than normal scan mode, however,
due to the mixture of pixel information, the camera resolution is low in this mode, and the Bayer pattern CCD
camera loses color information.
Figure 7.
2x2 Binning