Chapter 5
– Setup Menu
68
8.
The
Custom Spectral Calibration
is complete.
9.
To use these new factors for subsequent measurements, make sure that
Custom Spectral Correction
is
Enabled
and that
One Time Calibration
is
Disabled
. The settings will rest after each measurement.
10.
To restore the Target instrument so that it uses only the factory calibration factors, make sure
Custom Spectral Correction
is set to
Disabled
.
A
S A
.TXT
F
ILE
Data from other instruments can be used as the baseline or
Source
values for the Custom Spectral Calibration.
The following criteria must be met:
A. The
Source
must have the same starting and ending wavelength points, and the same wavelength
(data) increment.
B. The file must have a
.TXT
file extension.
C. The file must have the following file format:
Line 1
– Title (Alpha / Numeric characters only – no spaces)
Line 2
– 380 (Starting Wavelength)
Line 3
– 780 (for PR-730/740/788) or 1080 (for PR-735/745) – Ending Wavelength
Line 4
– 1 (for PR-730/740/788) or 2 (for PR-735/745) – Wavelength (Data) Increment (in
nanometers)
Line 5 to Line N
– Spectral values
Sample
.txt
File:
M1
3.002
7.002
1.000
3.722937e-003
3.729762e-003
3.611207e-003
3.971064e-003
4.476557e-003
4.434512e-003
4.972667e-003
4.636224e-003
5.055876e-003
.
.
.
.
.
.
1.
Create and save the
Source
spectral values on the SD card. Make sure this file does not have the same
name as another file on the same card with a
.txt
extension.
2.
Insert the SD card in the
Target
instrument.
3.
Make sure
Custom Spectral Correction
is
Disabled
and that
One Time Calibration
is
Enabled
.
4.
Enter the name of the file from Step 1 by touching the Spectral File field to bring up the Spectral File Dialog.
Summary of Contents for Photo Research SpectraScan PR-7 Series
Page 2: ......
Page 10: ...PR 7XX User s Manual vi...
Page 32: ......
Page 84: ......
Page 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Page 92: ......
Page 102: ......
Page 108: ......
Page 117: ...Chapter 12 Connectivity 113...
Page 118: ......