![Jadak Photo Research SpectraScan PR-7 Series User Manual Download Page 53](http://html1.mh-extra.com/html/jadak/photo-research-spectrascan-pr-7-series/photo-research-spectrascan-pr-7-series_user-manual_3186341053.webp)
Chapter 5
– Setup Menu
49
H
OW TO
M
AKE
L*
U
*
V
*
OR
L*
A
*
B
*
M
EASUREMENTS
1.
Touch the ▼ or ▲ icons in the
Measurement Mode
field to select
L*u*v*
or
L*a*b*
.
2.
To make
L*u*v*
or
L*a*b*
measurements, it is necessary to first select a white reference (
Illuminant
).
Touch
Select
at the bottom of the Illuminant column. A screen similar to the following ap
pears:
FIGURE 40 - L*U*V* SETUP SCREEN FIGURE 41 - FILE RECALL SCREEN
3.
To access factory stored illuminants (Standard Illuminants)
, touch the ▼ or ▲ icons in the
Current View
field until
Standard Illuminants appears.
An
Illuminant
may also be a previously measured AND
STORED file
– for example a florescent lamp. If it is desirable to use a previous measurement, then
select
Internal RAM
or
External SD Card
(if an SD card is inserted).
OR
Press the
MEASURE
button then touch
Back
then
Recall
to accept the measured data.
FIGURE 42 - STANDARD ILLUMINANTS SCREEN
4.
If using a pre-stored illuminant, at the top of the screen that pops up, the title of the
Illuminant
appears
.
Touch
Next
(or
Previous
) to scroll through available choices.
5.
Touch
Back.
6.
Touch
Recall
to select the displayed illuminant. The
Standard Illuminant
screen updates to show the
data of the illuminant selected.
Summary of Contents for Photo Research SpectraScan PR-7 Series
Page 2: ......
Page 10: ...PR 7XX User s Manual vi...
Page 32: ......
Page 84: ......
Page 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Page 92: ......
Page 102: ......
Page 108: ......
Page 117: ...Chapter 12 Connectivity 113...
Page 118: ......