PR-655/670 User Manual
45
→
→
I
MPORTING
D
ATA
A
S A
.MEA
F
ILE
Although it would be unlikely to have to perform this procedure for two PR-655 / PR-670 instruments, it is
possible to save a measurement on the SD card of the
Source
instrument, and use that measurement data in the
Target
instrument.
1.
Create a new file on the SD card in the
Source
instrument. Make sure this file does not have the same name
as another file on the same card with a
.txt
extension. The software will always try to use files with a
.txt
extension first during the spectral calibration sequence.
2.
Measure the sample with the
Source
instrument.
3.
Save the measurement to the file from Step 1
. Make sure that this measurement is the only measurement
in the file
. If multiple measurements are stored in the file, then the most recent measurement (last
measurement saved) will be used to generate spectral factors in the
Target
spectroradiometer.
4.
Remove the SD card from the
Source
instrument and insert it into the
Target
instrument.
5.
Make sure
Custom Spectral Correction
is
Disabled
and that
One Time Calibration
is
Enabled
.
6.