PR-655/670 User Manual
132
412
Syntax:
D412, ffffffff.eee
where: ffffffff
= filename of stored measurement file. DOS file naming convention.
eee =
filename extension
Output Format
: qqqqq,dt,tm CRLF (Directory of stored Measurements in file).
where
:
qqqqqq
= ID of measurement
dt
= Date
tm
= Time
Output Example
:
601
Output Format
: qqqqq,<Primary Lens>, <Addon1>, <Addon2>, <Addon3>, <Aperture>,
<Units>, <Exposure Mode>,<Exposure Time (see note) > ,<Gain>,
<Cycles>, <CIE Observer>, DarkMode>, <SyncMode>, <CaptureMode>,
<SyncPeriod>CRLF
Note
:
The exposure time reported in 601 and 602 (following) reports 0 if the
instrument is set in Adaptive Exposure mode. To view the exposure time for
the last measurement when the instrument is in Adaptive Exposure mode,
send the D13 command.
Output Example
:
00000,0,-1,-1,-1,0,0,0,0,0,1,2,0,0,0,60.00
602
Output Format
: Current set report with text labels.
Dark mode values: for reports [601] and [602]
0
Disable Smart Dark
1
Enable Smart Dark
Output Example
:
00000,MS-75,None,None,None,1 deg,English,Adaptive,0 msec,Normal,1 cycles,2 deg,No
Smart Dark,No Sync,Standard Sensitivity,60.00 Hertz