PR-655/670 User Manual
119
Setup
Command
Description
SC
Select Add-on Accessory 3
An Add-on accessory is one that is used in conjunction with a primary accessory. For
example, a neutral density filter (Add-on Accessory) used with the MS-75 (Primary
Accessory). Up to 3 Add-on accessories can be specified for a measurement.
Syntax
: SCn[CR]
Where
: n = Accessory code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Accessory Codes can be found by running report 116 (command D116). See the Data
Codes section for specific details.
SD
Select Dark Current Mode (PR-670 only)
Two dark current modes are available
–
Standard and Smart Dark. In Standard Mode, the
instrument measures the detector dark current after each light measurement.
If Smart Dark is enabled and two successive measurements yield the same exposure time
then the dark current values from the first measurement are used for the second (and
possibly successive) measurements.
Syntax
: SDn[CR]
Where
: n=Dark Current Mode
0 = Disable Smart Dark
1 = Enable Smart Dark
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)