1.
Navigate to the Timed Measurements screen by touching Menu, then Setup then Timed Measurements.
2.
If it is desirable to repeat the timed measurement sequence, set
Continuous Measurements
to
Enabled.
3.
Set the time between measurements by touching the
Measurement Interval
filed. The following data
entry screen appears:
FIGURE 80 - MEASUREMENT INTERVAL ENTRY FIELD
4.
Enter the measurement interval (range is 1 to 86400 seconds [24 hours]) then press
Back
.
5.
Press the
Measurement Button
to initiate the Timed Measurement Sequence.
C
ONTINUOUS
M
EASUREMENTS
1.
Navigate to the
Timed Measurements
screen by touching
Menu,
then
Setup
then
Timed
Measurements.
2.
Touch the
Timed Measurements
field to that it displays
Enabled.
3.
Touch the
Continuous Measurements
filed until
Enabled
appears.
4.
Press the
Measure button
to initiate a continuous measurement.