PR-655/670 User Manual
122
Setup
Command
Description
SO
CIE Observer
Photometric and Colorimetric values can be calculated using either CIE 2
or 10
Standard
Observer data sets. Use this specifier to choose the CIE data set for calculations for the
next measurement. The default is 2
.
Syntax
: SOn[CR]
Where
: n = CIE Observer
2 = 2
10 = 10
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
SP
Primary Accessory
A Primary Accessory is one that replaces the standard objective lens (typically the MS-75)
during use and can be used in conjunction with an Add-on Accessory.
Syntax
: SPnn[CR]
Where
: nn = Accessory Code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Accessory Codes can be found by running report 116 (command D116). See the Data
Codes section for specific details.