4.14 Setting measurement conditions
● Set the measurement conditions.
(1) While referring to Table 4-5, move the editing cursor to the Edit Box of the measurement condition to be
entered in Excited Conditions and press
to confirm the selection.
(2) For those with a Popup List that opens, move the editing cursor to a suitable item and press
to
confirm the settings.
(3) When the Popup Edit Box opens, use the ten keys to enter numeric characters directly and press
,
or move the editing cursor to the numeric character to be changed and press
, or turn
to
change the number. Press
to confirm the settings.
Table 4-5 Measurement Conditions
Items
Selectable items/Input rules
Descriptions
Function
Sine
Only the sine wave is used for the single sheet testing mode.
Calc.
Normal
Selects a calculation mode.
μ
Frequency
10 Hz to 20 kHz
Set the frequency of an excitation signal.
Fixed Parameter
(Target parameter)
Bm (Max. magnetic
flux density)
0.001 to
99999
Set a target parameter value for measurement.
Hm
(Max. magnetic
field)
I1m (Max. excitation
current)
V2m (Max. induced
voltage)
Tolerance
0.1 to 100%
When Yoke
Comp. is
Disabled
Finish the testing when the measured value of
the target parameter reaches within
X
[%] of this
set value. Set the tolerance
X
.
When Yoke
Comp. is
Enabled
Proceed from the 1st measurement to the 2nd
measurement when the measured value of the
target parameter reaches within
X
[%] of this set
value. The 3rd measurement is completed when
Bm of the 3rd measurement reaches the Bm of
the 2nd measurement and falls within
X
[%] of
this set value.
*
Therefore, the target parameter of the 1st
measurement will be within the set Tolerance
range, but when testing continues until the 3rd
measurement,
the target parameter may not
fall within the Tolerance
, due to the Yoke
Comp. function.
Yoke Comp.
Disabled
Disables Yoke Comp.
Enabled
Enables Yoke Comp.
Summary of Contents for SY-956
Page 1: ...Instruction Manual Mini Single Sheet Tester SY 956...
Page 2: ...2014_2015 IWATSU TEST INSTRUMENTS CORPORATION All rights reserved...
Page 37: ...27 Fig 4 19 1st Measurement Screen Fig 4 20 3rd Measurement Screen...
Page 41: ...31 6 2 Outside appearance diagram Unit mm...
Page 42: ...32 6 3 Outside appearance of the fixture Unit mm...
Page 44: ...SY 956...