10
In conclusion, by substituting H
p
of formula (12) into the first term of formula (7),
(13),
the actual iron loss of only the single sheet P
p
can be calculated.
Yoke Comp. comes into effect when Yoke Comp. is Enabled, as explained in
4.15. When Yoke Comp. is
Disabled, testing is completed after the 1st measurement.
Still,
Yoke Comp. does not necessarily work at all times
. In order to make the magnetic flux density B
ya
of
the double-yoke in the 3rd measurement equal to the magnetic flux density B
ya
of the measuring yoke in the
2nd measurement, this instrument controls the amplitude of the signal generator precisely and adjusts the
excitation current i
1b
sent through the double-yoke. However, the generator cannot be controlled when this B
ya
is extremely small. When a warning message is displayed and the 3rd measurement does not proceed after the
2nd measurement, it is usually the case that this B
ya
is extremely small (i.e., measurement conditions are such
that the core loss related to the measuring yoke is extremely small).
Fig. 2-4 Summary of the measurement principles
P
p
=
1
T
∫ H
p
dB
p
dt
dt
T
0
Signal
generator
Shunt resistor
Relay 1
B Coil
Single sheet
Excitation
coil
Measuring yoke
Double-yoke
Relay 2
Relay 3
B Coil
B Coil
Excitation
coil
Summary of Contents for SY-956
Page 1: ...Instruction Manual Mini Single Sheet Tester SY 956...
Page 2: ...2014_2015 IWATSU TEST INSTRUMENTS CORPORATION All rights reserved...
Page 37: ...27 Fig 4 19 1st Measurement Screen Fig 4 20 3rd Measurement Screen...
Page 41: ...31 6 2 Outside appearance diagram Unit mm...
Page 42: ...32 6 3 Outside appearance of the fixture Unit mm...
Page 44: ...SY 956...