7
IQ Power Fantom
5201292 Rev A
The relay contacts are rated for a maximum of 1 amp at 24 volts, with a resistive
load.
Figure 2. User Interface Relay Output (Typical Indicator Schematic)
Figure 3. User Interface Relay Output (Typical Light Tree Schematic)
In a typical indicator or light tree application:
• The green light indicates the unit is operating correctly or in standby.
• The red light indicates the unit in fault or not powered.
Figure 4. User Interface Relay Output (Typical PLC Schematic)
In a typical PLC application:
• The PLC input will be held high during normal operation or in standby.
• The PLC input will be low on fault or if the unit is not powered.
Inputs
Two inputs are available at the User Interface. The automatic point cleaner may
be activated with a momentary application of voltage to terminal 5. The unit can
be placed and held in standby mode with the application of a voltage to terminal 6.