6. MAINTENANCE
Ionizing Emitter Points
1. Dust or dirt on the ionization points will reduce effectiveness of the blower.
Ionization points must be cleaned periodically to prevent deposits from
accumulating.
2. Simply turn OFF and back ON. The automatic point cleaner will cycle three
times to clean the ionizing points. If the automatic point cleaner should
become inoperative, there is a Point Cleaner button adjacent to the ionized air
outlet. Push the button several times to clean the ion emitter points.
3. Frequency of cleaning should be at least once monthly, or as determined by
inspections based on operating conditions.
Air Intake Filter
1. Filter cleaning or replacement frequency depends upon the cleanliness of the
operating environment. Inspect filters weekly, clean or replace as required.
2. Turn the unit OFF and remove air filter from intake.
3. The air filter is open cell polyurethane. It may be cleaned with mild soap and
water. Rinse thoroughly and allow to dry before reinstalling.
4. To install air filter, place over intake grille. Tuck edges of air filter in recess
around perimeter of intake grille.
Ionizer Module
CAUTION
–
Risk of Injury
Turn off and disconnect unit from line voltage before removing ionizer
module. Use care handling ionizer module, puncture hazard, sharp pins
are present.
MISE EN GARDE – Risque de blessure
Éteindre et débrancher l’unité de la tension secteur avant de retirer le
module ioniseur. Manipuler le module ioniseur avec précaution, risque de
piqûre, broches pointues présentes.
1. Turn unit OFF.
2. Loosen (4) screws located at corners of ionized air outlet. The screws are captive
and will remain with ionizing module.
3. Remove ionizing module. The side opposite the point cleaner button contains
electrical connectors. It may be necessary to grasp this side and rock the
ionizing module gently to disconnect the connectors.