3.8.1.3.2.
.
PH SENSOR CALIBRATION
PRESS ENT. WAIT
PH CALIBRATION
SAMPLE PH = 7
ENT
PRES
TIME
EC
Q
PUMP
PH
Introduce the
Sensor in the Buffer
of pH7, press ENT
and wait
WAIT
0%
100%
PRESS ENT. WAIT
PH CALIBRATION
SAMPLE PH = 4
Introduce the
Sensor in the Buffer
pH4, and press ENT
and wait
WAIT
0%
100%
CALIBRATION SUCCES
PRESS ENT
Validate the pH Calibration by pressing ENTER.
Q
EC
PH
250
1.05
7.6
A
F02
5.4bar
B
CAL
PROG
ALM
CONT
HIST
AUX
mS
m3h
ENT
PRES
TIME
PUMP
PH
EC
Q
3. CONFIGURATION
Q
EC
PH
250
1.05
7.6
A
F02
5.4bar
B
1
4
EC 31%
PH 86%
mS
m3h
Q
EC
PH
250
1.05
7.6
A
F02
5.4bar
B
CAL
PROG
ALM
CONT
HIST
AUX
mS
m3h
ENT
PRES
TIME
PUMP
PH
EC
Q
PRESS ENT. WAIT
EC CALIBRATION
SAMPLE EC= 0.00
ENT
PRES
TIME
Q
PUMP
PH
Leave the Sensor on
the air, press ENT and
wait..
WAIT
0%
100%
PRESS ENT. WAIT
PH CALIBRATION
SAMPLE EC = 1. 40
Introduce the Sensor in the
Buffer of 1.40mS, press
ENT and wait.
WAIT
0%
100%
CALIBRATION SUCCES
PRESS ENT
Validate the EC Calibration by pressing ENTER.
EC
3.8.1.3.3 EC SENSOR CALIBRATION.
Q
EC
PH
250
1.05
7.6
A
F02
5.4bar
B
1
4
EC 31%
PH 86%
mS
m3h
WATER CONTROLLER 3000
USER'S MANUAL
3.8.1.4 PRESSURE TRANSMITTER CALIBRATION
TIME
Q
PUMP
PH
EC
PRES
20 mA 4.0
PRESS
Units:
bar
4 mA 0.0
ENT
Change Units by pr/-. Press > to scroll on the Menu and introduce the Pressure corresponding to 4mA and
20mA. Validate values pressing ENT.
PRES
TIME
PUMP
PH
EC
Q
17